Logo image
A swept-field multi-channel aspiration condenser for low-ppm level detection
Conference paper

A swept-field multi-channel aspiration condenser for low-ppm level detection

Chien-Fu Hsueh and Kea-Tiong Tang
2014 IEEE International Symposium on Bioelectronics and Bioinformatics, IEEE ISBB 2014, 6820895
2014

Abstract

aspiration condenser Ion mobility spectrometry swept-field
IMS(Ion Mobility Spectrometry) is a technique has been widely used for hazardous compounds detection. IMS can operates in atmospheric condition with low detection limit and fast response time. It became popular in various application, such as detection of chemical warfare agent, toxic industrial compounds and explosives. In this paper, we present a low-cost, easy to assembly device as an ion mobility spectrometry. This device is based on the theory of the first order differential aspiration condenser described by Puumalainen. Unlike the original device proposed by Puumalainen, the electric field is swept in discrete steps and can be control by user for specific application. Tammet transform is introduced to our system for expanding gas recognition ability. The result shows our system is capable for low-ppm level gas detection. © 2014 IEEE.