A voxel-based partial volume correction in nuclear medicine
Conference paper

A voxel-based partial volume correction in nuclear medicine

Hsuan-Ming Huang, Ing-Tsung Hsiao, Christian Wietholt and Ching-Han Hsu
Progress in Biomedical Optics and Imaging - Proceedings of SPIE, Vol.6144 III, 61446P
2006

Abstract

Nuclear medicine Partial Volume Correction Restoration and deblurring
Most partial volume correction (PVC) methods are ROI-based, and assume uniform activity within each ROI. Here, we extended a PVC method, developed by Roussel et al (JNM, 1998) called geometric transfer matrix (GTM), to a voxel-based PVC approach called v-GTM which accounts non-uniform activity within each ROI. The v-GTM method was evaluated using simulated data (perfect co-registered MRIs). We investigated the influence of noise, the effect of compensating detector response during iterative reconstruction methods and the effect of non-uniform activity. For simulated data, noise did not affect the accuracy of v-GTM method seriously. When detector response compensation was applied in iterative reconstruction, both PVC methods did not improve the recovery values. In the non-uniform experiment, v-GTM had slightly better recovery values and less bias than those of GTM. Conclusion: v-GTM resulted better recovery values, and might be useful for PVC in small regions of interest.