Design of an x-ray 3D microtomography beamline for full-field hard x-ray projection imaging at the Taiwan photon source
Conference paper   Open access

Design of an x-ray 3D microtomography beamline for full-field hard x-ray projection imaging at the Taiwan photon source

Shih-Hung Chang, Chao-Yu Chang, Chin-Yen Liu, Chia-Feng Chang, Liang-Chih Chiang, Bo-Yi Liao, Yu-Sheng Chen, Hsiang-Hsin Chen, Shun-Min Yang, Tsung-Tse Lee, …
AIP Conference Proceedings, Vol.2054, 060042
01/2019

Abstract

Physics and Astronomy (all)
The X-ray beamline TPS 02A is making use of the bending magnet source at the Taiwan Photon Source (TPS), which is dedicated to performing full-field hard X-ray projection imaging at micron size spatial resolutions in the photon energy range of 5 - 25 keV. It provides the photon flux >4.2 × 10 <sup>11</sup> photons·s <sup>-1</sup> with energy resolution <1.72 × 10-3, the beam size is 35.5 (H) × 8.5 (V) mm <sup>2</sup> and beam divergence is 1.2 (H) × 0.3 (V) mrad <sup>2</sup> at the sample position. The microscope is capable of examining anatomical information and internal distribution of microstructures by tuning the energy of the photon source to the absorption edge of the samples through the use of a monochromator. Phase contrast tomography is also available for imaging the fine details of in the morphology of the sample.