Logo image
Determination of entrance skin doses and organ doses for medical X ray examinations
Conference paper   Peer reviewed

Determination of entrance skin doses and organ doses for medical X ray examinations

C.J. Tung, C.Y. Cheng, T.C. Chao, H.Y. Tsai and H.Y. Tsai
Radiation Protection Dosimetry, Vol.85(1-4), pp.417-420
1999

Abstract

A national survey of patient doses for diagnostic X ray radiographs is planned in Taiwan. Entrance skin doses and organ doses for all installed X ray machines will be investigated. A pilot study has been carried out for the national survey to develop a protocol for the dose assessment. Entrance skin doses and organ doses were measured by thermoluminescence dosemeters and calculated by Monte Carlo simulations for several X ray examinations. The conversion factor from free air entrance absorbed dose to entrance skin dose was derived. A formula for the computation of entrance skin doses from inputs of kV(p), mA.s, source to skin distance, aluminium filtration, and generator rectifying was constructed. Organ doses were measured using a RANDO phantom and calculated using a mathematical phantom. All data will be passed to the Atomic Energy Council for developing a programme of national survey and regulatory controls for diagnostic X ray examinations.

Metrics

1 Record Views

Details

Logo image