QUANTITATIVE POINT-OF-CARE CADMIUM DETECTION USING A MEMS SPECTROMETER-BASED LFA READER
會議論文集

QUANTITATIVE POINT-OF-CARE CADMIUM DETECTION USING A MEMS SPECTROMETER-BASED LFA READER

Yu-Chen Hsu, Weileun Fang, Chitsung Hong, Cheng-Hao Ko, Jyun-Wei Wen 和 Guan-Yi Lin
Proceedings, IEEE micro electro mechanical systems, 頁碼.538-541
Proceedings IEEE Micro Electro Mechanical Systems
01/01/2026
Web of Science ID: WOS:001799185400134

摘要

Engineering Engineering, Electrical & Electronic Engineering, Mechanical Nanoscience & Nanotechnology Science & Technology Science & Technology - Other Topics Technology
We present a portable sensor for quantitative detection of the critical heavy metal toxin, Cadmium (Cd), by LFA, replacing high Coefficient of Variation (CV) image-based analysis with a MEMS spectrometer. Fabricated via X-ray lithography, the spectral chip achieves a 200:1 high aspect ratio (HAR) and 5 nm wavelength resolution at VIS similar to NIR bandwidth. This direct spectroscopic system, over 75% smaller than benchtop units, enhances total measurement precision to 13% CV (including strip variability), while the instrument's standalone CV is under 3%. The sensor achieved a 2.75ppb Limit of Detection (LOD) with a calibration curve R-2 of 0.966 across a 4 similar to 56ppb range.