利用游標尺延遲線技術實現的快速單斜率類比數位轉換器
Thesis

利用游標尺延遲線技術實現的快速單斜率類比數位轉換器

林蔚峰
Masters, 國立清華大學, 工程與系統科學系
2009

Abstract

單斜率類比數位轉換器 游標尺延遲線 時間數位轉換器 延遲鎖相迴路 Single Slope ADC Vernier Delay Line Time-to-Digital Converter Delay Lock Loop
Integrating type analog-to-digital converter (ADC) is commonly used for nuclear radiation spectrometers for pulse height measurement. Vernier delay line (VDL) techniques have been proposed for time measurement and can reduce conversion time significantly. With improving time resolution of VDLs, the timing error due to comparator delays and the non-linearity of ramp generator become a concern for improving the performance of the single slope ADC. The present work is to use a two-level VDL for high resolution timing measurement and to develop an amplitude-to-time converter with high precision to reduce timing errors in the time conversion process. A 9-bit single slope ADC is realized with the process of TSMC CMOS 0.18um 1P6M. Its sample rate is 5Msps. And the minimum time resolution is about 150ps, the DNL is within -0.4~+0.5 LSB, and the INL is within -0.1~1.1 LSB.