熱發光劑量照野電子污染誤差函原子學核子學科學
ATOMICSNUCLEAR-SCIENCESCIENCE
以熱發光劑量計度量10MVX 射束內不同照野電子污染與給予劑量比值。7×7厘米2 照野電子污梁比值為4.0 %,6×6厘米2 為3.3% ,5×5厘米2 為2.5 %,4×4厘 米2為1.7%,3×3厘米2 為0.9 %,污梁電子能量為1.94Mev.其在側向成誤差函數 分佈。 #2811148