Logo image
A 1.72μW, 23.2fj/conversion step successive approach ADC for bio-medical signal acquisition.
Journal article

A 1.72μW, 23.2fj/conversion step successive approach ADC for bio-medical signal acquisition.

Wei-Sheng Wang and Kea-Tiong Tang
Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Conference, Vol.2011, pp.2703-2706
2011

Abstract

This paper presents a successive approach register (SAR) analog-to-digital converter (ADC) with a novel hybrid digital-to-analog converter (DAC) architecture: half junction splitting (J.S.) and half binary weighted capacitor DAC. This DAC maintains low power consumption of J.S. DAC and the high signal-to-noise plus distortion ratio (SNDR) of binary weighted capacitor DAC. The power dissipation of the circuit is 1.72 μW, SNDR 59.17 dB, spurious free dynamic range (SFDR) 73.39 dB, and the FOM 23.2 fj/conversion step with 0.9 V supply voltage. The proposed circuit is fabricated in TSMC 0.18 μm 1P6M CMOS process technology.

Metrics

1 Record Views

Details

Logo image