A parameter study of pencil beam proton dose distributions for the treatment of ocular melanoma utilizing spot scanning
期刊文章   同儕審查

A parameter study of pencil beam proton dose distributions for the treatment of ocular melanoma utilizing spot scanning

Kenneth Sutherland, Satoshi Miyajima, Hiroyuki Date, Hiroki Shirato, Masayori Ishikawa, Masao Murakami, Mitsuru Yamagiwa, Paul Bolton, Toshiki Tajima 和 Fang-Hsin Chen
Radiological physics and technology, 卷.3(1), 頁碼.16-22
01/01/2010
PMID: 20821097
Web of Science ID: WOS:000219369300003

摘要

Life Sciences & Biomedicine Radiology, Nuclear Medicine & Medical Imaging Science & Technology
The results of Monte Carlo calculated dose distributions of proton treatment of ocular melanoma are presented. An efficient spot scanning method utilizing active energy modulation, which also minimizes the number of target spots was developed. We simulated various parameter values for the particle energy spread and the pencil beam diameter in order to determine values suitable for medical treatment. We found that a 2.5-mm-diameter proton beam with a 5% Gaussian energy spread was suitable for treatment of ocular melanoma while preserving vision for the typical case that we simulated. The energy spectra and the required proton current were also calculated and are reported. The results are intended to serve as a guideline for a new class of low-cost, compact accelerators.