Ambient neutron dose equivalent during proton therapy using wobbling scanning system: Measurements and calculations
期刊文章

Ambient neutron dose equivalent during proton therapy using wobbling scanning system: Measurements and calculations

Yung-Chieh Lin, Chung-Chi Lee, Tsi-Chian ChaoHui-Yu Tsai
Radiation Physics and Chemistry, 卷.140, 頁碼.290-294
11/2017

摘要

H*(10)/D Monte Carlo Proton therapy Secondary neutron WENDI-II Wobbling scatter nozzle Radiation
Neutron production is a concern in proton therapy, particularly in scattering proton beam delivery systems. Despite this fact, little is known about the effects of secondary neutron exposure around wobbling scattered proton treatment nozzles. The objective of this study was to estimate the neutron dose level resulting from the use of a wobbling scattered proton treatment unit. We applied the Monte Carlo method for predict the ambient neutron dose equivalent, H*(10), per absorbed dose at the treatment isocenter, D, in the proton therapy center of Chang Gung Memorial Hospital, Linkou, Taiwan. For a 190-MeV proton beam, H * (10)/D values typically decreased with the distance from the isocenter, being 1.106 mSv/Gy at the isocenter versus 0.112 mSv/Gy at a distance of 150 cm from the isocenter. The H * (10)/D values generally decreased as the neutron receptors moved away from the isocenter, and increased when the angle from the initial beam axis increased. The ambient neutron dose equivalents were observed to be slightly lower in the direction of multileaf collimator movement. For radiation protection, the central axis of a proton-treated patient is suggested to be at the 0° angle of the beam. If the beam direction at the 90° angle is necessary, the patient axis is suggested to be along with the direction of MLC movement. Our study provides the neutron dose level and neutron energy fluence for the first wobbling proton system at the proton therapy center of Chang Gung Memorial Hospital.

相關連結

指標

1 檢視次數

詳細資料

Logo image