Artifact reduction of ultrasound Nakagami imaging by combining multifocus image reconstruction and the noise-assisted correlation algorithm
期刊文章   同儕審查

Artifact reduction of ultrasound Nakagami imaging by combining multifocus image reconstruction and the noise-assisted correlation algorithm

Po-Hsiang TsuiYu-Wei Tsai
Ultrasonic imaging, 卷.37(1), 頁碼.53-69
01/01/2015
PMID: 24626567
Web of Science ID: WOS:000350487200005

摘要

Algorithms Artifacts Humans Image Enhancement - methods Image Processing, Computer-Assisted - methods Models, Statistical Phantoms, Imaging Scattering, Radiation Signal Processing, Computer-Assisted Transducers Ultrasonography - methods
Several studies have investigated Nakagami imaging to complement the B-scan in tissue characterization. The noise-induced artifact and the parameter ambiguity effect can affect performance of Nakagami imaging in the detection of variations in scatterer concentration. This study combined multifocus image reconstruction and the noise-assisted correlation algorithm (NCA) into the algorithm of Nakagami imaging to suppress the artifacts. A single-element imaging system equipped with a 5 MHz transducer was used to perform the brightness/depth (B/D) scanning of agar phantoms with scatterer concentrations ranging from 2 to 32 scatterers/mm(3). Experiments were also carried out on a mass with some strong point reflectors in a breast phantom using a commercial scanner with a 7.5 MHz linear array transducer operated at multifocus mode. The multifocus radiofrequency (RF) signals after the NCA process were used for Nakagami imaging. In the experiments on agar phantoms, an increasing scatterer concentration from 2 to 32 scatterers/mm(3) led to backscattered statistics ranging from pre-Rayleigh to Rayleigh distributions, corresponding to the increase in the Nakagami parameter measured in the focal zone from 0.1 to 0.8. However, the artifacts in the far field resulted in the Nakagami parameters of various scatterer concentrations to be close to 1 (Rayleigh distribution), making Nakagami imaging difficult to characterize scatterers. In the same scatterer concentration range, multifocus Nakagami imaging with the NCA simultaneously suppressed two types of artifacts, making the Nakagami parameter increase from 0.1 to 0.8 in the focal zone and from 0.18 to 0.7 in the far field, respectively. In the breast phantom experiments, the backscattered statistics of the mass corresponded to a high degree of pre-Rayleigh distribution. The Nakagami parameter of the mass before and after artifact reduction was 0.7 and 0.37, respectively. The results demonstrated that the proposed method for artifact reduction allows a sensitive and effective scatterer characterization by Nakagami imaging.

相關連結

詳細資料

Logo image