CMOS 256-Pixel/480-Pixel Photovoltaic-Powered Subretinal Prosthetic Chips with Wide Image Dynamic Range and Bi/Four-Directional Sharing Electrodes and Their Ex Vivo Experimental Validations with Mice
期刊文章   同儕審查

CMOS 256-Pixel/480-Pixel Photovoltaic-Powered Subretinal Prosthetic Chips with Wide Image Dynamic Range and Bi/Four-Directional Sharing Electrodes and Their Ex Vivo Experimental Validations with Mice

Chung-Yu Wu, Chi-Kuan Tseng, Jung-Hsing Liao, Chuan-Chin Chiao, Fang-Liang Chu, Yueh-Chun Tsai, Jun OhtaToshihiko Noda
IEEE Transactions on Circuits and Systems I: Regular Papers, 卷.67(10), 頁碼.3273-3283
10/2020

摘要

CMOS image sensor Divisional power supply scheme implantable chip subretinal prostheses Electrical and Electronic Engineering
CMOS photovoltaic-powered 256-pixel/480-pixel implantable chips with a wide image dynamic range and shared electrodes are proposed and fabricated in CIS technology for subretinal prostheses. The infra-red (IR) light is incident only on photovoltaic cells of the chip whereas the visible light is mainly incident on pixels. The proposed adaptive background cancellation circuit (ABCC) is adopted to increase the image dynamic range so that the subretinal chips can adapt for different surrounding illuminances. A two-dimensional equivalent electrode model for electric crosstalk simulation is proposed to analyze the crosstalk charges and enable the stimulation pattern design with minimum crosstalk for both chips. Moreover, the bi-directional sharing electrodes (BDSEs) are used in the 256-pixel chip to increase electrode size under the same chip area and boost the stimulation charges to 11.4 nC. The four-directional sharing electrodes (FDSEs) are used to share the electrodes in both vertical and horizontal directions and achieve 480 pixels with nearly the same electrode size under the same pixel array area. The maximum stimulation charge is 9.8 nC. The dynamic range of image light is increased to 31.65 dB (24.9 dB) in the fabricated 256-pixel (480-pixel) chip. The functions of the chip have been successfully validated by both electrical measurement and ex vivo patch clamp experiments with the retinas of mice.

相關連結

指標

1 檢視次數

詳細資料

Logo image