Logo image
Collimation-enhanced micro-radiography in real-time
期刊文章   同儕審查

Collimation-enhanced micro-radiography in real-time

Y. Hwu, W.-L. Tsai, H.-H. Hsieh, Jung Ho Je, Hye-Suk Kang, In-Woo Kim, Kyu-Ho Lee, Hee-Joung Kim, B. LaiG. Margaritondo
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 卷.467-468(PART II), 頁碼.1294-1300
2001

摘要

Nuclear and High Energy Physics Instrumentation
Highly collimated synchrotron sources are the essential ingredient of novel radiology and micro-radiology techniques for materials science, medicine and biology-with contrast based on the refractive index rather than on absorption. We were able to implement a real-time version of such techniques. Successful test studies included thin-foil breaking and electrodeposition processes in the areas concerning materials properties. Meanwhile, extensive effort was dedicated to obtain high-quality X-ray images of live specimens on a microscopic scale and in real time. These tests achieved a lateral resolution better than 2 μm and a time per image of 10 ms, with clear potential for further improvements. The images revealed in great detail moving parts of different specimens, opening the way to a wide variety of applications in the life sciences, medicine and materials science. © 2001 Published by Elsevier Science B.V.