Logo image
Detection of C-reactive protein on an integrated microfluidic system by utilizing field-effect transistors and aptamers
期刊文章   同儕審查

Detection of C-reactive protein on an integrated microfluidic system by utilizing field-effect transistors and aptamers

Wei-Chieh Kao, Yen-Wen Chen, Chia-Ho Chu, Wen-Hsin Chang, Shu-Chu Shiesh, Yu-Lin WangGwo-Bin Lee
Biomicrofluidics, 卷.11(4), 044105
07/2017

摘要

Molecular Biology Materials Science (all) Genetics Condensed Matter Physics Physical and Theoretical Chemistry
Cardiovascular diseases (CVDs) cause more than 17 × 10 6 deaths worldwide on a yearly basis. Early diagnosis of CVDs is therefore of great need. The C-reactive protein (CRP) is an important biomarker for analyzing the risks of CVDs. In this work, CRP-specific aptamers with high sensitivity and specificity and fieldeffect- transistor (FET) devices were used to recognize and detect CRP by using an integrated microfluidic system automatically while consuming less volumes of reagents and samples (about 5 μm). In order to package the FET device into the microfluidic chip, a new method to prevent liquid leakage was proposed. Sensitive detection of CRP has been demonstrated on the developed microfluidic system. It is the first time that aptamer-FET assays could be realized on an integrated microfluidic system. Experimental results showed that the aptamer-FET assay was capable of detecting CRP with concentrations ranging from 0.625 mg/l to 10.000 mg/l, which may be promising for early diagnosis of CVDs.

相關連結

指標

1 檢視次數

詳細資料

Logo image