Development and Clinical Validation of a Sensitive Lateral Flow Assay for Rapid Urine Fentanyl Screening in the Emergency Department
期刊文章   開放取用(OA)

Development and Clinical Validation of a Sensitive Lateral Flow Assay for Rapid Urine Fentanyl Screening in the Emergency Department

Zhao Li, Hui Chen, Sheng Feng, Kengku Liu 和 Ping Wang
Clinical chemistry, 卷.66(2), 頁碼.324-332
02/2020
PMID: 32040576

摘要

emergency room fentanyl lateral flow screening sensitivity Medicine (all)
BACKGROUND: Rapid identification of fentanyl at the point-of-care is critical. Urine fentanyl concentrations in overdose cases start at single-digit nanograms per milliliter. No fentanyl point-of-care assay with a cutoff at single-digit nanograms per milliliter is available. METHODS: A competitive lateral flow assay (LFA) was developed using gold nanoparticles and optimized for rapid screening of fentanyl in 5 minutes. Urine samples from 2 cohorts of emergency department (ED) patients were tested using the LFA and LC-MS/MS. The 2 cohorts consisted of 218 consecutive ED patients with urine drug-of-abuse screen orders and 7 ED patients with clinically suspected fentanyl overdose, respectively. RESULTS: The LFA detected fentanyl (≥1 ng/mL) and the major metabolite norfentanyl (≥10 ng/mL) with high precision. There was no cross-reactivity with amphetamine, cocaine, morphine, tetrahydrocannabinol, methadone, buprenorphine, naloxone, and acetaminophen at 1000 ng/mL and 0.03%, 0.4%, and 0.05% cross-reactivity with carfentanil, risperidone, and 9-hydroxyrisperidone, respectively. In 218 consecutive ED patients, the prevalence of cases with fentanyl ≥1 ng/mL or norfentanyl ≥10 ng/mL was 5.5%. The clinical sensitivity and specificity of the LFA were 100% (95% CI, 75.8-100%) and 99.5% (95% CI, 97.3-99.9%), respectively. The positive and negative predictive values were 92.3% (95% CI, 66.7-98.6%) and 100% (95% CI, 98.2-100%), respectively. The concordance between the LFA and LC-MS/MS was 100% in the 7 suspected fentanyl overdose cases (5 positive, 2 negative). CONCLUSIONS: The LFA can detect fentanyl and norfentanyl with high clinical sensitivity and specificity in the ED population with rapid fentanyl screening needs.

檔案與連結 (1)

url
https://doi.org/10.1093/clinchem/hvz023檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image