Logo image
Development of a desktop-sized thermal lens microscope
期刊文章   開放取用(OA)

Development of a desktop-sized thermal lens microscope

Shinichiro Hiki, Manabu Tokeshi, Akihide HibaraTakehiko Kitamori
Bunseki Kagaku, 卷.52(8), 頁碼.569-574
08/2003

摘要

Non-fluorescent molecule Thermal lens effect Thermal lens microscope Ultra high sensitive detection Analytical Chemistry
A desktop-sized thermal lens microscope (DT-TLM) having high sensitivity and wide applicability was developed. The operationality of the DT-TLM compared with that of a big-sized TLM has been remarkably improved by simplifying the optics and optimizing the optical configuration. In order to evaluate the performance, the DT-TLM was applied to the ultrasensitive detection of a non-fluorescent molecule (dye: Sunset Yellow) in water. The concentration dependence of the thermal lens signal (calibration curve) showed good linearity in the range of 1 × 10 -1 × 10 M. From the value of twice the standard deviation (2σ) of this calibration curve, the lower limit of the quantitative determination was estimated to be 1.10 × 10 M. The lower limit of detection was estimated as 8.27 × 10 M from the conditions of the signal-to-noise ratio S/N = 2. These obtained values were almost the same as that of the big-sized TLM.

檔案與連結 (1)

url
https://doi.org/10.2116/bunsekikagaku.52.569檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image