Development of a disposable force-sensing glove for clinicians and demonstration of its force measurements on patients during rehabilitation following anterior cruciate ligament reconstruction surgery
期刊文章   開放取用(OA)   同儕審查

Development of a disposable force-sensing glove for clinicians and demonstration of its force measurements on patients during rehabilitation following anterior cruciate ligament reconstruction surgery

Wei-Chih Wang, David R. LindersDavid J. Nuckley
Journal of Biomedical Optics, 卷.22(12), 127002
12/2017
PMID: 29264895

摘要

anterior cruciate ligament anterior cruciate ligament rehabilitation bendloss force feedback force sensor glove sensor haptics pain threshold patch sensor strength measurement Electronic Optical and Magnetic Materials Biomaterials Atomic and Molecular Physics and Optics Biomedical Engineering
For many clinicians, their effectiveness is dependent on the magnitude of forces they manually apply to their patients. However, current state-of-the-art care strategies lack quantitative feedback, making it difficult to provide consistent care over time and among multiple clinicians. To provide real-time quantitative feedback to clinicians, we have developed a disposable glove with a force sensor embedded in the fingertips or palm. The sensor is based on the fiber-optic bendloss effect whereby light intensity from an infrared source is attenuated as the fiber is bent between a series of corrugated teeth. The sensor fabricated has a very low profile (10×7×1 mm) and has demonstrated high sensitivity, accuracy, range, and durability. Force feedback up to 90 N with an average force threshold at 0.19 N and average sensor resolution at 0.05 N has been demonstrated. A preliminary clinical study has also been conducted with anterior cruciate ligament reconstruction patients who show significant range of motion improvement when treated with the force-sensing glove.

檔案與連結 (1)

url
https://doi.org/10.1117/1.JBO.22.12.127002檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image