Logo image
Difference in Tracer Delay-induced Effect among Deconvolution Algorithms in CT Perfusion Analysis: Quantitative Evaluation with Digital Phantoms
期刊文章   同儕審查

Difference in Tracer Delay-induced Effect among Deconvolution Algorithms in CT Perfusion Analysis: Quantitative Evaluation with Digital Phantoms

Kohsuke Kudo, Makoto Sasaki, Kuniaki Ogasawara, Satoshi Terae, Shigeru Ehara, Hiroki Shirato 和 Fang-Hsin Chen
Radiology, 卷.251(1), 頁碼.241-249
01/04/2009
PMID: 19190251
Web of Science ID: WOS:000265643000028

摘要

Life Sciences & Biomedicine Radiology, Nuclear Medicine & Medical Imaging Science & Technology
Institutional review board approval and informed consent were obtained. The purpose was to evaluate the differences in tracer delay-induced effects of various deconvolution algorithms for computed tomographic (CT) perfusion imaging by using digital phantoms created from actual source data. Three methods of singular value decomposition (SVD) were evaluated. For standard SVD (sSVD), the delays induced significant errors in cerebral blood flow and mean transit time. In contrast, for block-circulant SVD (bSVD), these values remained virtually unchanged, whereas for delay-corrected SVD (dSVD), mild changes were observed. bSVD was superior to sSVD and dSVD for avoiding the tracer delay-induced effects in CT perfusion imaging. (C) RSNA, 2009