Effect of geometric models on convergence rate in iterative PET image reconstructions
Journal article   Open access   Peer reviewed

Effect of geometric models on convergence rate in iterative PET image reconstructions

C.-C. Liu, C.-H. Hsu, I.-T. Hsiao and K.M. Lin
Journal of Instrumentation, Vol.4(5), P05010
2009

Abstract

Gamma camera, SPECT, PET PET/CT,coronary CT angiography (CTA) Image reconstruction in medical imaging
Iterative PET image reconstructions can improve quantitation accuracy by explicitly modeling photon-limited nature and physical effects of coincident photons. Geometric model in iterative reconstructions defines the mapping between image and sinogram domains based on scanner's geometry, which affects the accuracy of image results and the convergence rate of reconstruction algorithms. This paper examines the convergence rates of a reconstruction algorithm with three PET geometric models: interpolative, area-based, and solid-angle. The iterative algorithm used in this study is the maximum likelihood expectation-maximization (MLEM) algorithm. Experimental data are generated by the GATE package which simulates the Inveon microPET system. The comparison of convergence rate is based on the plot of log-likelihood value versus iteration number. From the plots of log-likelihood curves, the results from solid-angle model consistently reach the highest values at early iterations. It means that the MLEM algorithm with the solid-angle model will converge faster than the other two models. The experimental results indicate that the solid-angle model is a favorable geometric model for faster iterative PET image reconstruction. © 2009 IOP Publishing Ltd and SISSA.
url
https://doi.org/10.1088/1748-0221/4/05/P05010View
Published (Version of record) Open

Related links

Metrics

1 Record Views

Details

Logo image