Environmental monitoring using a HPGe-NaI(Tl) compton suppression spectrometer
Journal article   Peer reviewed

Environmental monitoring using a HPGe-NaI(Tl) compton suppression spectrometer

CHIEN CHUNG and Cheng-Jong Lee,
Nuclear Inst. and Methods in Physics Research, A, Vol.273(1), pp.436-440
01/12/1988

Abstract

A multipurpose Compton suppression γ-ray spectrometer is used for environmental monitoring. A high-purity germanium detector is positioned in a large size NaI(Tl) guard detector and coupled to an anticoincidence gating system. Environmental samples of rainwater collected after the Chernobyl accident are monitored with and without anti-Compton operation. The optimum ratio of the peak-to-Compton edge at 662 keV γ-rays is 472:1. The detection limit of 137 Cs concentration in a pretreated rain sample is 0.09 Bq for a 20000 s counting period. Characteristics of this spectrometer and features of the γ-ray spectrum collected for fallout monitoring are discussed. © 1988.