Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial
期刊文章

Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial

C.-H. Liao, D.A. Spain, C.-C. Chen, C.-T. Cheng, W.-C. Lin, D.-R. Ho, H.-F. Lin 和 F. Catena
World Journal of Emergency Surgery, 卷.20(1)
2025
Web of Science ID: WOS:001406182300001

摘要

Abdominal compartment syndrome Capsular sensor Digital health Intraabdominal pressure Adult Aged Feasibility Studies Female Humans Intra-Abdominal Hypertension Laparoscopy Male Middle Aged Monitoring, Physiologic Pilot Projects Pressure Prospective Studies Reproducibility of Results adult aged devices diagnosis feasibility study female human intraabdominal hypertension laparoscopy male middle aged pathophysiology physiologic monitoring pilot study pressure procedures prospective study reproducibility
Background: Intrabdominal pressure (IAP) is an important parameter. Elevated IAP can reduce visceral perfusion, lead to intraabdominal hypertension, and result in life-threatening abdominal compartment syndrome. While ingestible capsular devices have been used for various abdominal diagnoses, their application in continuous IAP monitoring remains unproven. Method: We conducted a prospective clinical trial to evaluate the feasibility of IAP measurement using a digital capsule PressureDOT, an ingestible capsule equipped with wireless transmission capability and a pressure sensor, then compared its reliability with conventional intravesical method. Patients undergoing laparoscopic or robotic surgeries were recruited. During surgery, we created pneumoperitoneum by inflating CO2 into the peritoneal cavity and IAP was simultaneously monitored using both the ingestible capsules and intravesical measurements from Foley catheter. We assessed the feasibility of signal transmission and the accuracy of pressure measurements. Results: Six patients were enrolled in this pilot study. No adverse events were reported, and the average first-intake time was within 24 h. All capsules were successfully expelled, with an average excretion time of 81 h. In the summarized data, the mean IAPdot is 0.6 mmHg lower than the IAPivp, with a standard deviation of 1.68 mmHg. However, capsule measurements showed excellent correlation with intravesical IAP measurements, with an intraclass correlation coefficient of 0.916 (95% CI: 0.8821–0.9320). Conclusion: Our study demonstrates the feasibility and safety of using digital capsules for continuous IAP monitoring, providing the agreement between IAP measurements from digital capsules and conventional intravesical measurement within a near-normal pressure. © The Author(s) 2024.

檔案與連結 (2)

url
https://www.scopus.com/inward/record.uri?eid=2-s2.0-85217090169&doi=10.1186%2fs13017-024-00569-0&partnerID=40&md5=47dd9551031970a1f1b60134fad27431檢視
url
https://doi.org/10.1186/s13017-024-00569-0檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image