Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor
期刊文章   同儕審查

Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor

Hiroki Shirato, Shinichi Shimizu, Kei Kitamura, Takeshi Nishioka, Kenji Kagei, Seiko Hashimoto, Hidefumi Aoyama, Tatsuya Kunieda, Nobuo Shinohara, Hirotoshi Dosaka-Akita, …
International journal of radiation oncology, biology, physics, 卷.48(2), 頁碼.435-442
01/09/2000
PMID: 10974459
Web of Science ID: WOS:000089069100019

摘要

Conformal radiotherapy Fluoroscopy Localization Pattern recognition Radiotherapy
Purpose: To achieve precise three-dimensional (3D) conformal radiotherapy for mobile tumors, a new radiotherapy system and its treatment planning system were developed and used for clinical practice. Methods and Materials: We developed a linear accelerator synchronized with a fluoroscopic real-time tumor tracking system by which 3D coordinates of a 2.0-mm gold marker in the tumor can be determined every 0.03 second. The 3D relationships between the marker and the tumor at different respiratory phases are evaluated using CT image at each respiratory phase, whereby the optimum phase can be selected to synchronize with irradiation (4D treatment planning). The linac is triggered to irradiate the tumor only when the marker is located within the region of the planned coordinates relative to the isocenter. Results: The coordinates of the marker were detected with an accuracy of ± 1 mm during radiotherapy in the phantom experiment. The time delay between recognition of the marker position and the start or stop of megavoltage X-ray irradiation was 0.03 second. Fourteen patients with various tumors were treated by conformal radiotherapy with a “tight” planning target volume (PTV) margin. They were surviving without relapse or complications with a median follow-up of 6 months. Conclusion: Fluoroscopic real-time tumor tracking radiotherapy following 4D treatment planning was developed and shown to be feasible to improve the accuracy of the radiotherapy for mobile tumors.

相關連結

指標

172 readers on Mendeley
1 readers on CiteULike

詳細資料

Logo image