Logo image
Influence of respiration on dose calculation in stereotactic body radiotherapy of the lung
期刊文章   同儕審查

Influence of respiration on dose calculation in stereotactic body radiotherapy of the lung

Rie Yamazaki, Rikiya Onimaru, Norio Katoh, Tetsuya Inoue, Takeshi Nishioka, Hiroki Shirato, Hiroyuki Date 和 Fang-Hsin Chen
Radiological physics and technology, 卷.7(2), 頁碼.284-289
01/07/2014
PMID: 24643842
Web of Science ID: WOS:000219416000012

摘要

Life Sciences & Biomedicine Radiology, Nuclear Medicine & Medical Imaging Science & Technology
Our purpose in this study was to evaluate the variation in calculated doses caused by respiration in stereotactic body radiotherapy (SBRT) of the lung. The study targeted ten patients who underwent SBRT for lung tumors. CT images were acquired during free breathing and in the inhalation and exhalation phases. We compared the CT image at inhalation with the image at exhalation so as to measure the change in lung volume, variation in the CT value, and displacement of the chest wall. The lung volume change was shown to be correlated with the maximum of the chest wall motion and with the variation in the CT value. A statistically significant difference was observed in the CT values between inhalation and exhalation (p < 0.05). The total dose variation at the isocenter was confined within +/-2 %. However, the dose from individual beams can vary significantly when the chest wall moves more than 10 mm in natural breathing.