Logo image
Integration of flow injection analysis and zeptomole-level detection of the Fe(II)-o-phenanthroline complex
期刊文章   開放取用(OA)   同儕審查

Integration of flow injection analysis and zeptomole-level detection of the Fe(II)-o-phenanthroline complex

Kiyoshi Sato, Manabu Tokeshi, Takehiko KitamoriTsuguo Sawada
Analytical Sciences, 卷.15(7), 頁碼.641-645
1999

摘要

Flow injection analysis Molecular diffusion On-chip integration Photothermal microscope Analytical Chemistry
Microchannels having a 150x100 μm cross section were fabricated in a quartz glass chip as a component in an integrated flow injection analysis (FIA) system. They were put to use for flow, mixing, reaction, and detection. The reaction system was a chelating reaction of divalent iron ion with o- phenanthroline (o-phen), and a photothermal microscope was applied for the ultra-sensitive detection of the non-fluorescent reaction product. Nano liter levels of solutions were introduced and transported by capillary action and mixed by molecular diffusion. Zeptomole levels of the reaction product were detected quantitatively. This was the first demonstration of an on-chip chemical determination device which integrates the primitive FIA system without using electroosmotic liquid control or fluorometric determination.

檔案與連結 (1)

url
https://doi.org/10.2116/analsci.15.641檢視