Intrafractional Baseline Shift or Drift of Lung Tumor Motion During Gated Radiation Therapy With a Real-Time Tumor-Tracking System
期刊文章   同儕審查

Intrafractional Baseline Shift or Drift of Lung Tumor Motion During Gated Radiation Therapy With a Real-Time Tumor-Tracking System

Seishin Takao, Naoki Miyamoto, Taeko Matsuura, Rikiya Onimaru, Norio Katoh, Tetsuya Inoue, Kenneth Lee Sutherland, Ryusuke Suzuki, Hiroki Shirato, Shinichi Shimizu, …
International journal of radiation oncology, biology, physics, 卷.94(1), 頁碼.172-180
01/01/2016
PMID: 26700711
Web of Science ID: WOS:000366574700038

摘要

Carcinoma, Non-Small-Cell Lung - pathology Carcinoma, Non-Small-Cell Lung - surgery Computer Systems Dose Fractionation Fiducial Markers Humans Lung Neoplasms - pathology Lung Neoplasms - surgery Movement Patient Positioning - instrumentation Patient Positioning - statistics & numerical data Radiosurgery - instrumentation Radiosurgery - methods Radiotherapy Setup Errors - prevention & control Respiration Time Factors
To investigate the frequency and amplitude of baseline shift or drift (shift/drift) of lung tumors in stereotactic body radiation therapy (SBRT), using a real-time tumor-tracking radiation therapy (RTRT) system. Sixty-eight patients with peripheral lung tumors were treated with SBRT using the RTRT system. One of the fiducial markers implanted near the tumor was used for the real-time monitoring of the intrafractional tumor motion every 0.033 seconds by the RTRT system. When baseline shift/drift is determined by the system, the position of the treatment couch is adjusted to compensate for the shift/drift. Therefore, the changes in the couch position correspond to the baseline shift/drift in the tumor motion. The frequency and amount of adjustment to the couch positions in the left-right (LR), cranio-caudal (CC), and antero-posterior (AP) directions have been analyzed for 335 fractions administered to 68 patients. The average change in position of the treatment couch during the treatment time was 0.45 ± 2.23 mm (mean ± standard deviation), -1.65 ± 5.95 mm, and 1.50 ± 2.54 mm in the LR, CC, and AP directions, respectively. Overall the baseline shift/drift occurs toward the cranial and posterior directions. The incidence of baseline shift/drift exceeding 3 mm was 6.0%, 15.5%, 14.0%, and 42.1% for the LR, CC, AP, and for the square-root of sum of 3 directions, respectively, within 10 minutes of the start of treatment, and 23.0%, 37.6%, 32.5%, and 71.6% within 30 minutes. Real-time monitoring and frequent adjustments of the couch position and/or adding appropriate margins are suggested to be essential to compensate for possible underdosages due to baseline shift/drift in SBRT for lung cancers.

相關連結

詳細資料

Logo image