On-chip antibody immobilization for on-demand and rapid immunoassay on a microfluidic chip
期刊文章   開放取用(OA)   同儕審查

On-chip antibody immobilization for on-demand and rapid immunoassay on a microfluidic chip

Toshinori Ohashi, Kazuma MawatariTakehiko Kitamori
Biomicrofluidics, 卷.4(3), 032207
2010

摘要

Molecular Biology Materials Science (all) Genetics Condensed Matter Physics Physical and Theoretical Chemistry
Immunoassay is one of the important applications of microfluidic chips and many methodologies were reported for decreasing sample/reagent volume, shortening assay time, and so on. Micro-enzyme-linked immunosorbent assay (micro-ELISA) is our method that utilizes packed microbeads in the microfluidic channel and the immunoreactions are induced on the beads surface. Due to the large surface-tovolume ratio and small analytical volume, excellent performances have been verified in assay time and sample/reagent volume. In order to realize the micro-ELISA, one of the important processes is the immobilization of antibody on the beads surface. Previously, the immobilization process was performed in a macroscale tube by physisorption of antibody, and long time (2 h) and large amount of antibody (or high concentration) were required for the immobilization. In addition, the processes including the reaction and washing were laborious, and changing the analyte was not easy. In this research, we integrated the immobilization process into a microfluidic chip by applying the avidin-biotin surface chemistry. The integration enabled very fast (1 min) immobilization with very small amount of precious antibody consumption (100 ng) for one assay. Because the laborious immobilization process can be automatically performed on the microfluidic chip, ELISA method became very easy. On-demand immunoassay was also possible just by changing the antibodies without using large amount of precious antibodies. Finally, the analytical performance was investigated by measuring C-reactive protein and good performance (limit of detection <20 ng/ml) was verified. © 2010 American Institute of Physics.

檔案與連結 (1)

url
https://doi.org/10.1063/1.3437592檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image