On-chip connector valve for immunoaffinity chromatography in a microfluidic chip
期刊文章   同儕審查

On-chip connector valve for immunoaffinity chromatography in a microfluidic chip

Kiyohito Shimura, Yutaka Koyama, Kae SatoTakehiko Kitamori
Journal of Separation Science, 卷.30(10), 頁碼.1477-1481
07/2007
PMID: 17623428

摘要

Chromatography Immunoaffinity Microchip Microfluidics Valve Analytical Chemistry Filtration and Separation
A miniature valve that operates between a chip port and a tube fitting was developed. The valve functions by means of a rotor, 3 mm in diameter and 1.5 mm in height, made of Teflon, with a 0.2-mm diameter hole at its center that is co-axial with the tube fitting. It also has a radial groove, 0.85 mm. long, 0.2 mm wide, and 0.2 mm deep, at the bottom surface, starting at its center. The chip port and the tube fitting have an offset of 0.75 mm., and, thus, the rotation of the rotor can make an on and off connection between the chip port and the groove, which is connected to the tubing. The valve had a pressure resistance of at least 1.0 MPa. The on-chip valve can be placed in position by adding only a single part, a valve rotor, and no changes in the fabrication of the glass microchip are required. Since the valve functions as a part of a connector, we refer to it as an on-chip connector valve. Immunoaffinity chromatography of a fluorescence-labeled recombinant antibody fragment was carried out in a glass microchip using the valves. © 2007 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim.