P value estimation for a dead time distorted Poisson distribution measured using a nuclear radiation counting system
Journal article   Peer reviewed

P value estimation for a dead time distorted Poisson distribution measured using a nuclear radiation counting system

Chih-Hao Lee and Ceming Wang
Journal of Radioanalytical and Nuclear Chemistry, Vol.303(3), pp.1659-1662
2015

Abstract

Chi-square test Detector dead time Nonparalyzable and paralyzable P value
The P value used to estimate the statistical significance of a nuclear counting system was evaluated in the context of dead time. As the detector dead time increases, the P value also increases larger than 0.5. For a number of samples with batch size of 50, the dead time of a radiation detection system should be kept within 23 % if the P value is to be less than 0.9.