Logo image
Performances of gamma ray measurement using bismuth germanate detector after thermal neutron bombardment
Journal article

Performances of gamma ray measurement using bismuth germanate detector after thermal neutron bombardment

Cheng-Jong Lee and CHIEN CHUNG
International Journal of Radiation Applications and Instrumentation. Part, Vol.42(8), pp.729-733
1991

Abstract

Performance of a 3.8 × 3.8 cm BGO detector used as a gamma ray spectrometer was tested after thermal neutron bombardment. The drift of photopeak channel as well as change of absolute counting efficiency and photopeak energy resolution became severe as the BGO detector was bombarded with thermal neutrons to 10 13 n th /cm 2 . Irradiation with thermal neutrons beyond 10 15 n th /cm 2 disabled the BGO detector as a gamma ray spectrometer due to the intense interference from internal radioactivity induced by neutron reactions. © 1991.