Tung-Yuan Hsiao, Lu-Kai Wang, Tzung-Yuang Chen, Ching-Fang Yu, Cheng-Ya Pan, Chun-Chieh Wang, Chien-Yu Lin, I-Chun Cho, Fang-Hsin Chen, Huan Niu, …
IEEE Transactions on Radiation and Plasma Medical Sciences, 卷.8(1), 頁碼.88-94
01/2024

摘要

Animals;Biology;Hospitals;Monitoring;Particle beams;Protons;Radiation effects Atomic and Molecular Physics and Optics Instrumentation Radiology Nuclear Medicine and Imaging

The purpose of this study was to create a platform for conducting small animal proton FLASH irradiation at a research-specific therapy center using an auxiliary proton beam line. A dose monitoring system was implemented to provide real-time irradiation dose and dose rate time structure information, utilizing a transmission ionization chamber, a plastic scintillator-based beam position monitor, and a Poor Man Faraday Cup. Both the TIC and PMFC were equipped with a custom-made fast-reading current integral electronics device, allowing for real-time dose monitoring with a resolution of 0.5 ms. Experimental testing showed that the maximum dose that could be delivered to the target in a single pulse of 230 MeV protons was approximately 20 Gy, during a pulse period of 199 ms, at a dose rate of 100 Gy/s. The establishment of a dedicated proton research platform at Chang Gung Memorial Hospital has enabled investigation of the biological effects of FLASH irradiation on small animals. The final setup data serves as a useful reference for both planning future experiments and improving the facility for beam users. This study&null achievement provides a valuable foundation for further research on proton FLASH irradiation, potentially offering new insights into its applications for cancer treatment.

相關連結

指標

1 檢視次數

詳細資料

Logo image