Simple dose rate measurements for a very high synchrotron X-ray flux
期刊文章   開放取用(OA)   同儕審查

Simple dose rate measurements for a very high synchrotron X-ray flux

Chi-Jen Liu, Chang-Hai Wang, Cheng-Liang Wang, Y. Hwu, Chien-Yi LinG. Margaritondo
Journal of Synchrotron Radiation, 卷.16(3), 頁碼.395-397
2009
PMID: 19395805

摘要

Bleaching Dosimetry Fricke dosimetry Intense X-rays Methylene blue Radiation Nuclear and High Energy Physics Instrumentation
Dose measurements based on methylene blue (MB) bleaching, widely used for ultraviolet light, can also be applied to X-rays including very high flux levels. This method has been tested by using both MB bleaching and Fricke dosimetry for a conventional monochromatic X-ray source and then for 'white-beam' synchrotron radiation. The results show that MB bleaching dosimetry can easily measure X-ray doses up to at least 10 5 Gy s -1 , as long as the MB concentration is sufficiently high. This condition can be verified from the deviations from linearity of the bleaching versus exposure time. © 2009 International Union of Crystallography.