Single-Channel Bioimpedance Measurement for Wearable Continuous Blood Pressure Monitoring
期刊文章   開放取用(OA)   同儕審查

Single-Channel Bioimpedance Measurement for Wearable Continuous Blood Pressure Monitoring

Ting-Wei Wang, Wen-Xiang Chen, Hsiao-Wei ChuShien-Fong Lin
IEEE Transactions on Instrumentation and Measurement, 卷.70, 9247300
2021

摘要

Continuous blood pressure (BP) measurement impedance plethysmography (IPG) wearable device Instrumentation Electrical and Electronic Engineering
Pulse transit time (PTT) sensing technology is rapidly evolving for cuffless continuous blood pressure (BP) monitoring. However, PTT-based BP methods must implement multidevices that could impede wearable applications. This work presents the study and development of a single-channel wrist-worn impedance plethysmography (IPG) system for wearable continuous BP measurement. IPG-based BP mathematical model with two-step calibration was derived for single physiological signal acquisition. The high-resolution IPG measurement was performed by optimized injection current frequency and electrode arrangement in response to obvious impedance difference between systolic BP (SBP) and diastolic BP (DBP) feature points in the IPG waveform, resulting in an impedance difference of 594 mΩ. The BP accuracy results within 30 subjects indicated that SBP estimation error was 2.01 ± 1.40 mmHg and 1.83 ± 1.29% in terms of mean absolute error (MAE) and mean absolute percentage error (MAPE), and 2.26 ± 1.43 mmHg and 4.96 ± 3.02% for DBP estimation. The novel system could potentially be implemented as a wearable cardiovascular monitoring device with two-in-one functions (continuous BP and heart rate monitoring) using a single physiological sensor in the future.

檔案與連結 (1)

url
https://doi.org/10.1109/TIM.2020.3035578檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image