Three-dimensional intrafractional motion of breast during tangential breast irradiation monitored with high-sampling frequency using a real-time tumor-tracking radiotherapy system
期刊文章   同儕審查

Three-dimensional intrafractional motion of breast during tangential breast irradiation monitored with high-sampling frequency using a real-time tumor-tracking radiotherapy system

Rumiko Kinoshita, Shinichi Shimizu, Hiroshi Taguchi, Norio Katoh, Masaharu Fujino, Rikiya Onimaru, Hidefumi Aoyama, Fumi Katoh, Tokuhiko Omatsu, Masayori Ishikawa, …
International journal of radiation oncology, biology, physics, 卷.70(3), 頁碼.931-934
01/03/2008
PMID: 18164868
Web of Science ID: WOS:000253224000043

摘要

Life Sciences & Biomedicine Oncology Radiology, Nuclear Medicine & Medical Imaging Science & Technology
Purpose: To evaluate the three-dimensional intrafraction motion of the breast during tangential breast irradiation using a real-time tracking radiotherapy (RT) system with a high-sampling frequency. Methods and Materials: A total of 17 patients with breast cancer who had received breast conservation RT were included in this study. A 2.0-mm gold marker was placed on the skin near the nipple of the breast for RT. A fluoroscopic real-time tumor-tracking RT system was used to monitor the marker. The range of motion of each patient was calculated in three directions. Results: The mean standard deviation of the range of respiratory motion was 1.0 +/- 0.6 mm (median, 0.9; 95% confidence interval [CI] of the marker position, 0.4-2.6), 1.3 +/- 0.5 mm (median, 1.1; 95 % CI, 0.5-2.5), and 2.6 +/- 1.4 (median, 2.3; 95 % CI, 1.0-6.9) for the right-left, craniocaudal, and anteroposterior direction, respectively. No correlation was found between the range of motion and the body mass index or respiratory function. The mean +/- standard deviation of the absolute value of the baseline shift in the right-left, craniocaudal, and anteroposterior direction was 0.2 +/- 0.2 mm (range, 0.0-0.8 mm), 0.3 +/- 0.2 mm (range, 0.0-0.7 mm), and 0.8 +/- 0.7 mm (range, 0.1-1.8 mm), respectively. Conclusion: Both the range of motion and the baseline shift were within a few millimeters in each direction. As long as the conventional wedge-pair technique and the proper immobilization are used, the intrafraction three-dimensional change in the breast surface did not much influence the dose distribution. (C) 2008 Elsevier Inc.

相關連結

詳細資料

Logo image