Use of Implanted Markers and Interportal Adjustment With Real-Time Tracking Radiotherapy System to Reduce Intrafraction Prostate Motion
期刊文章   同儕審查

Use of Implanted Markers and Interportal Adjustment With Real-Time Tracking Radiotherapy System to Reduce Intrafraction Prostate Motion

Shinichi Shimizu, Yasuhiro Osaka, Nobuo Shinohara, Ataru Sazawa, Kentaro Nishioka, Ryusuke Suzuki, Rikiya Onimaru, Hiroki Shirato 和 Fang-Hsin Chen
International journal of radiation oncology, biology, physics, 卷.81(4), 頁碼.e393-e399
15/11/2011
PMID: 21658857
Web of Science ID: WOS:000309412300025

摘要

Image-guided radiotherapy Intrafraction organ motion Prostate Radiotherapy
Interportal adjustment was applied to patients with prostate cancer using three fiducial markers and two sets of fluoroscopy in a real-time tumor-tracking radiotherapy (RTRT) system. The incidence of table position adjustment required to keep intrafractional uncertainty within 2.0 mm was investigated in this study. The coordinates of the center of gravity of the three fiducial markers were measured at the start of every portal irradiation in intensity-modulated radiotherapy (IMRT) with seven ports. The table position was adjusted to the planned position if the discrepancy was larger than 2.0 mm in the anterior–posterior (AP), cranial–caudal (CC), or left–right (LR) directions. In total, we analyzed 4,541 observations in 20 patients who received 70 Gy in 30 fractions (7.6 times a day on average). The incidence of table position adjustment at 10 minutes from the initial setup of each treatment was 14.2%, 12.3%, and 5.0% of the observations in the AP, CC, and LR directions, respectively. The accumulated incidence of the table position adjustment was significantly higher at 10 minutes than at 2 minutes for AP (p = 0.0033) and CC (p = 0.0110) but not LR (p = 0.4296). An adjustment greater than 5 mm was required at least once in the treatment period in 11 (55%) patients. Interportal adjustment of table position was required in more than 10% of portal irradiations during the 10-minute period after initial setup to maintain treatment accuracy within 2.0 mm.

相關連結

詳細資料

Logo image