Use of a thermal lens microscope in integrated catecholamine determination on a microchip
期刊文章

Use of a thermal lens microscope in integrated catecholamine determination on a microchip

Hussein M. Sorouraddin, Akihide HibaraTakehiko Kitamori
Analytical and Bioanalytical Chemistry, 卷.371(2), 頁碼.91-96
2001
PMID: 11678204

摘要

Analytical Chemistry Biochemistry
A new method for determination for catecholamines (CA) utilizing microchip technology and a thermal lens microscope has been developed. Microchannels with a 250 nmxlOO iim cross section were used for mixing, reaction, and detection. Epinephrine (EP), nor-epinephrine (NE), dopamine (DA), and L-dopa (LD) were determined by using coloring oxidization to aminochromes by sodium metaperiodate. A thermal lens microscope (TLM) was used for detection of the product. The sensitivity of the system was comparable for the four CA and required only 15 s for mixing of sample and reagent. The calibration lines indicated excellent linearity for concentrations of 5-20 μg mL . The relative standard deviations for 10 μg mL-1 solution were 1.08, 2.18, 2.2, and 2.5% for EP, NE, DA, and LD, respectively. CA in pharmaceutical injections were also determined by use of the system and the results correlated very well with nominal values. Results obtained by use of the integrated system suggested there was a sufficient possibility to realize high-throughput medical diagnosis systems. © Springer-Verlag 2001.