Validation of continuous intraabdominal pressure measurement: feasibility and accuracy assessment using a capsular device in in-vivo studies
期刊文章

Validation of continuous intraabdominal pressure measurement: feasibility and accuracy assessment using a capsular device in in-vivo studies

D.-R. Ho, C.-T. Cheng, C.-H. Ouyang, W.-C. Lin 和 C.-H. Liao
World Journal of Emergency Surgery, 卷.19(1)
2024
Web of Science ID: WOS:001255374000001

摘要

Abdominal compartment syndrome Capsular sensor Digital health Intraabdominal pressure Abdominal Cavity Animals Disease Models, Animal Feasibility Studies Intra-Abdominal Hypertension Monitoring, Physiologic Pressure Reproducibility of Results Swine abdominal cavity animal devices diagnosis disease model feasibility study intraabdominal hypertension pathophysiology physiologic monitoring physiology pig pressure procedures reproducibility
Background: Monitoring Intraabdominal Pressure (IAP) is essential in critical care, as elevated IAP can lead to severe complications, including Abdominal Compartment Syndrome (ACS). Advances in technology, such as digital capsules, have opened new avenues for measuring IAP non-invasively. This study assesses the feasibility and effectiveness of using a capsular device for IAP measurement in an animal model. Method: In our controlled experiment, we anesthetized pigs and simulated elevated IAP conditions by infusing CO2 into the peritoneal cavity. We compared IAP measurements obtained from three different methods: an intravesical catheter (IAPivp), a capsular device (IAPdot), and a direct peritoneal catheter (IAPdir). The data from these methods were analyzed to evaluate agreement and accuracy. Results: The capsular sensor (IAPdot) provided continuous and accurate detection of IAP over 144 h, with a total of 53,065,487 measurement triplets recorded. The correlation coefficient (R²) between IAPdot and IAPdir was excellent at 0.9241, demonstrating high agreement. Similarly, IAPivp and IAPdir showed strong correlation with an R² of 0.9168. Conclusion: The use of capsular sensors for continuous and accurate assessment of IAP marks a significant advancement in the field of critical care monitoring. The high correlation between measurements from different locations and methods underscores the potential of capsular devices to transform clinical practices by providing reliable, non-invasive IAP monitoring. © The Author(s) 2024.

檔案與連結 (2)

url
https://www.scopus.com/inward/record.uri?eid=2-s2.0-85197167415&doi=10.1186%2fs13017-024-00553-8&partnerID=40&md5=378aad8b34ba5bcc486db35747eddc4f檢視
url
https://doi.org/10.1186/s13017-024-00553-8檢視
已出版(紀錄版本) 開放

相關連結

指標

1 檢視次數

詳細資料

Logo image