電腦斷層掃描之自動管電流調控技術的劑量與影像品質
Journal article

電腦斷層掃描之自動管電流調控技術的劑量與影像品質

(Shu-Fan Tsai) 蔡書汎, (Yu-Shan Tyan) 田雨生, (Yi-Lun Hung) 洪儀倫, (Ying-Lan Liao) 廖英蘭 and (Hui-Yu Tsai) 蔡惠予
中華放射線技術學雜誌, Vol.34(2), pp.81-89
01/06/2010

Abstract

有效劑量 image quality 自動管電流調控技術 Automatic Tube Current Modulation Effective Dose Computed Tomography 影像品質 電腦斷層掃描機
Computed tomography (CT) has contributed about 50% of the collective dose from medical exposure. The dose from CT has been considered seriously. Thus, automatic tube current modulation is designed to reasonably lower radiation dose without affecting image quality. The purpose of this study is to evaluate how automatic tube current modulation to optimize radiation dose and to impact image quality. A standard reference anthropomorphic phantom took head, neck, chest, abdomen, pelvic, and whole abdomen (including pelvic) scanning on a multidetector CT scanner (Somatom Sensation 64, Siemens) with and without automatic tube current modulation (CARE Dose 4D, Siemens). Organ doses and effective doses were estimated by the high sensitivity and tissue-equivalent thermoluminescent dosimeters (TLD). Effective dose conversion factors were also used to fast calculate effective dose. Image quality was assessed by analyzing the noise level of images. The effective dose decreased 5~47% in head, chest, abdomen, pelvic and whol